Clinical trial sample storage is the controlled holding of processed samples at the condition the sponsor’s lab manual specifies, from the end of processing until shipment to the laboratory. Sound storage gives every aliquot a recorded location, a continuous temperature history and a named person for every move.
The test of that record often comes at night. In the fictional study ONC-214, the alarm on minus 80 °C freezer F3 sounds at 02:14 on a Saturday. The on-call technician moves the contents to a backup unit, and the monitoring system later shows that F3 peaked at minus 52 °C for 3 hours 40 minutes. On Monday the research team needs one list: which participants’ samples were in F3, in which boxes, and for how long. The freezer map was last updated three weeks ago.
This guide covers what sample storage involves, what a location record must hold, where storage breaks down, how to respond to a temperature excursion, and what monitors and inspectors check. It forms part of our guide to clinical trial sample management.
What Does Clinical Trial Sample Storage Involve?
Storage starts the moment an aliquot leaves the processing bench. It ends when the sample is packed for shipment, analysed locally or destroyed. The lab manual sets the condition for each sample type, and the site has to show that the condition held for every hour in between.
| Storage condition | Examples of sample types (the lab manual decides) | What the site records |
| Ambient (15 to 25 °C) | Whole blood for same-day shipment, some urine samples | Holding location and courier handover time, against the stability limit |
| 2 to 8 °C | Some safety samples awaiting the local laboratory, some swabs | Fridge identity, time placed and removed, minimum and maximum temperatures |
| Minus 20 °C | Some serum and urine aliquots | Freezer, box and position, time placed, continuous temperature record |
| Minus 80 °C | Pharmacokinetic (PK) plasma, biomarker serum and plasma aliquots | Freezer, rack, box and position, time placed, continuous monitoring with an alarm |
| Liquid nitrogen vapour phase (below minus 150 °C) | Viable cells such as peripheral blood mononuclear cells (PBMCs) | Dewar, canister and position, temperature and nitrogen level record |
| Dry ice (about minus 78 °C) | Frozen aliquots packed for shipment | Packing time, dry ice quantity, courier handover time |
Most studies use several rows of this table at once. A single oncology visit can produce safety bloods for the local NHS laboratory, PK plasma for minus 80 °C and cells for liquid nitrogen, each with its own storage evidence. Shared freezers then hold boxes from many studies side by side.
What Should a Sample Location Record Hold?
A location record answers one question for any sample at any time: where is it now, and where has it been? It works as an address, from the storage unit down to a single position in a box. Six elements make that address complete.
- Storage unit, a uniquely named freezer, fridge or dewar with its required range and monitoring probe, which links the sample to the temperature record for that unit.
- Shelf or rack, the physical division inside the unit, which narrows a search from the whole freezer to one rack.
- Box, labelled with its own identifier and study, which keeps studies apart inside shared freezers.
- Position, the grid reference inside the box (A1 to I9 in a 9 by 9 box), which makes each aliquot findable without opening every box.
- Time placed and person, the event time and the delegated member of staff, which shows the sample reached its storage condition within the lab manual limit.
- Moves, every transfer between units, including moves during defrosting, maintenance and alarms, which keeps the history unbroken.
The storage entry also closes the processing clock. The time an aliquot is placed in the freezer is the time the lab manual’s collection-to-freezing limit is measured against, as our guide to clinical trial sample tracking explains.
Where Does Sample Storage Break Down?
Storage records fail at the points where samples move and the record stays still. Five patterns recur at research sites:
- Batch-updated freezer maps, kept on paper or a spreadsheet and brought up to date at the end of the day or the week.
- Unrecorded moves during defrosting, maintenance or an alarm, noted on a sticky label or left out altogether.
- Shared freezers holding several studies, with boxes consolidated or relabelled over time.
- Disconnected temperature records, held by estates or a separate monitoring system with no link to the samples inside each unit.
- Out-of-hours responders who can move samples safely and have no access to the study record.
The worked example below is fictional. It follows the ONC-214 excursion through a site that keeps these records apart.
| Time | Event | Record used | What is missing |
| Sat 02:14 | F3 alarm at minus 70 °C and rising | Monitoring system alert to the on-call technician | Which studies are stored in F3 |
| Sat 02:40 | Contents moved to backup freezer F5 | Handwritten note on the F3 door | New box positions in F5 |
| Sat 05:54 | F3 back below minus 75 °C | Monitoring system | Peak and duration known, affected samples unknown |
| Mon 09:00 | Research team learns of the alarm | Email from estates | Two days lost before any sponsor is told |
| Mon 15:30 | Affected list built from the freezer map | Spreadsheet last updated three weeks earlier | Eleven recent aliquots absent from the list |
The freezer recovered in under four hours. The list of affected samples took most of a week.
Freezer Maps Kept Apart Compared With Locations Recorded on Each Sample
The central difference is where the location lives. A separate freezer map describes the freezer. A location recorded on the sample record describes the sample, and the freezer view is built from those records.
| Aspect | Freezer map kept apart | Location recorded on each sample |
| Timing | Updated after the work, often in batches | Recorded as the aliquot is placed |
| Moves | Noted on paper during alarms and defrosts | Each move recorded with person, time and new position |
| Excursion scope | Built by hand from the map and from memory | Every sample in the affected unit listed from the record |
| Shipment | Boxes searched and positions crossed off | Samples picked by position and removed from the location record |
| Shared freezers | Each study keeps its own map | One site view across studies, filtered by study |
| Inspection | Map, log and monitoring printouts compared side by side | Location history shown for any sample on request |
What Is a Temperature Excursion for Stored Samples?
A temperature excursion is any period when a storage unit runs outside the range the lab manual sets for the samples it holds. The alarm is the monitoring system’s signal. The excursion is the event itself, defined by its peak temperature and its duration.
The decision on sample use rests with the sponsor and the central laboratory, which hold the stability data for each analyte. The site’s role is to detect the excursion, protect the samples, define the scope and report the facts. Every excursion report needs five facts:
- Unit and required range: which freezer, fridge or dewar, and the condition its samples need.
- Start, end and duration: taken from the monitoring record, with the time the site was alerted.
- Peak temperature: the warmest point reached, with the probe or logger that recorded it.
- Affected samples: every aliquot in the unit during the excursion, by participant, timepoint, box and position.
- Actions taken: moves to backup storage, with times, people and new positions.
Investigational medicinal product follows the same logic under its own rules. See what a site should do after an IMP temperature excursion.
How Should a Site Respond to a Sample Storage Excursion?
A written response turns a night-time alarm into a controlled event. The seven steps below suit most research sites. The site’s own standard operating procedure (SOP) and each study’s lab manual set the detail.
- Protect the samples. Move them to a qualified backup unit at the right condition and record each move with its new position.
- Capture the temperature record for the full event, including peak and duration, and keep the original data.
- Identify every affected sample from the location record, across all studies stored in the unit.
- Place the affected samples on hold, which stops them being shipped or used before a decision.
- Notify each sponsor or central laboratory within the timeline its lab manual sets, with the five facts above.
- Record the decision for each sample: release for use, use with an annotation, or discard under the sponsor’s instruction.
- Assess the event under the site’s deviation process, and open a corrective and preventive action (CAPA) where the cause could recur.
Repeat excursions on one unit point to a system cause, such as an ageing compressor, an overfull freezer or a worn door seal. The fix belongs in the quality system. See managing deviations in clinical research and a CAPA example walked through step by step.
Also Read: Investigational Product Storage Records: A Compliance Checklist
What Do Monitors and Inspectors Check About Sample Storage?
Monitors and inspectors test storage from both ends: from the freezer to the record, and from the record to the freezer. Six questions come up repeatedly.
- Can the site show the current location of a sample the laboratory has not yet received?
- Does each storage unit have a continuous temperature record, with alarms and responses documented?
- Are monitoring probes and loggers calibrated, with certificates on file?
- Is there a qualified backup unit, and was it used during the last alarm?
- Were excursions reported to the sponsor, and is the decision recorded for each sample?
- Are any samples held beyond their approved retention, or after a participant withdrew consent?
The expectations behind these questions are set in published frameworks. ICH E6(R3) expects trial records that are attributable, complete and retrievable, and the MHRA sets out its inspection approach in its good clinical practice inspection guidance. Cellular samples are relevant material under the Human Tissue Act 2004, and the Human Tissue Authority’s code of practice for research explains when storage needs a licence and when material held for a specific approved research project is exempt.
Laboratory certificates, monitoring calibration records and excursion correspondence belong in the investigator site file, where a monitor expects to find them.
Also Read: MHRA GCP Inspection Common Findings
What Are the Risks of Weak Sample Storage Records?
The risks fall on the data, the participants and the site’s standing with sponsors.
- Samples with an unknown storage history may be excluded from analysis, and each lost timepoint is permanent.
- An incomplete excursion list can let an affected sample reach the laboratory unflagged.
- Hours spent searching boxes during an excursion come out of research nurse and technician time.
- Samples kept after consent withdrawal or past their retention period create a legal exposure under human tissue legislation.
- Sponsors notice repeated storage queries and may weigh them in feasibility for the next study.
Storage problems usually start one step earlier, with kit supplies and processing times. See lab kit management in clinical trials for the supply side.
How Does AQ Labs Support Clinical Trial Sample Storage?
AQ Labs is the lab sample management module of AQ Platform. It records storage as part of each sample’s own history, on the same record as the site file.
- Each aliquot is recorded to a freezer, rack, box and position, which links every storage location record to the sample it holds.
- Storage positions are recorded by scan as the work happens, which replaces the separately maintained freezer map.
- A temperature excursion recorded for a storage unit identifies every sample stored in it and places them on hold, which stops affected samples being shipped or used before a decision.
- The sponsor’s decision is recorded for each sample, and releasing a sample after an excursion requires re-authentication by the person taking that action.
- Shipments accept only samples for the receiving laboratory and storage condition, and ship-day and stability checks run before packing.
AQ Labs works alongside temperature monitoring systems. The monitoring system keeps the readings and alarms, and AQ Labs links each excursion to the samples in the affected unit. AQ Labs is priced and deployed as its own module. Book a live demo to see an excursion traced from one freezer to every affected sample.
Frequently Asked Questions
What is clinical trial sample storage?
Clinical trial sample storage is the controlled holding of processed samples at the condition the sponsor’s lab manual specifies, from the end of processing until shipment, local analysis or destruction. Each aliquot needs a recorded location, a continuous temperature history and a named person for every move.
At what temperature are clinical trial samples stored?
The lab manual sets the condition for each sample type. Common conditions are ambient, 2 to 8 °C, minus 20 °C, minus 80 °C for most frozen plasma and serum aliquots, and liquid nitrogen vapour phase for viable cells.
What is a freezer map?
A freezer map is the record of which samples sit in which rack, box and position inside a storage unit. It is most reliable when built from a location recorded on each sample at the time of placement, with every later move recorded.
Who decides whether samples can be used after a temperature excursion?
The sponsor decides, usually with the central laboratory, because they hold the stability data for each analyte. The site protects the samples, identifies every affected aliquot, places them on hold, reports the facts and records the decision for each sample.
Does a research site need a Human Tissue Authority licence to store trial samples?
It depends on the material and the approvals. Storage of relevant material, which includes cellular samples, for research generally needs a Human Tissue Authority licence in England, Wales and Northern Ireland. Material held for a specific research project approved by a recognised research ethics committee can be exempt. Scotland has separate legislation.